Table 3:

Algorithm’s Diagnostic Performance for Individual Vessel Segments

No.
occlusions
(%)
No. controls
AUC
(95% CI)
Target Sens
(%)
(95%CI in %)
Target Spec
(%)
(95%CI in %)
Threshold
(range)
‘COLOR’
Jmax Sensitivity
(%)
(95%CI in %)
Jmax Specificity
(%)
(95%CI in %)
Jmax
(95%CI)
(threshold range)
‘COLOR’
Intracranial
ICA
incl. terminal
ICA
133
(20.03)
531
0.977
(0.965–0.989)
129/133
(96.99)
92.5–99.2
459/531
(86.44)
83.2–89.2
<60%–45%
YELLOW
118/133
(88.72)
82.1–93.5
521/531
(98.12)
96.6–99.1
0.8684
0.8174–0.9153
<45%
RED
M1-MCA 290
(35.32)
531
0.962
(0.948–0.976)
281/290
(96.90)
94.2–98.6
423/531
(79.66)
76.0–83.0
<75%–60%
GREEN
243/290
(83.79)
79.0–87.8
521/531
(98.12)
96.6–99.1
0.8191
0.7724–0.8494
<45%
RED
M2-MCA 60
(10.15)
531
0.874
(0.826–0.921)
54/60
(90.00)*
79.5–96.2
398/531
(74.95)
71.0–78.6
<80%–75%
BLUE
52/60
(86.67)
75.4–94.1
423/531
(79.66)
76.0–83.0
0.6633
0.5479–0.7327
<75%–60%
GREEN

Note. – ‘Target Sens’ is associated with ‘Threshold’ that yields ≥95% sensitivity at the highest specificity (‘Target Spec’).

Jmax is the maximum Youden index (sensitivity + specifity − 1) across all thresholds.

The DeLong algorithm was used to compute the Standard Error of the AUC.

*

For M2-MCA segments the ≥95% sensitivity target could not be met.