Skip to main content
. 2019 Nov 22;57(12):e01179-19. doi: 10.1128/JCM.01179-19

TABLE 1.

Reproducibility of the IgG capture assay

Sample Mean S/CO (SD)a CV%b
Intra-assay variability
    Plasma
        HTLV-1 sample A 21.84 (0.69) 3.1%
        HTLV-2 sample B 3.05 (0.09) 2.9%
    Oral fluid
        Pool with low S/CO 5.55 (0.15) 2.65%
        Pool with high S/CO 20.91 (0.84) 4%
Interassay variability
    Plasma
        HTLV-1 sample A 21.63 (1.39) 6.4%
        HTLV-1 sample B 7.25 (0.64) 8.9%
        HTLV-1 sample C 21.27 (0.93) 4.4%
        HTLV-1 sample D 19.73 (1.17) 5.9%
        HTLV-1 sample E 8.73 (0.94) 10.8%
        HTLV-1 sample F 6.39 (0.33) 5.2%
        HTLV-1 sample G 5.35 (0.35) 6.5%
        HTLV-2 sample A 2.96 (0.03) 1.1%
    Oral fluid
        Pool with low S/CO 6.05 (0.53) 8.7%
        Pool with high S/CO 22.36 (1.5) 6.7%
a

S/CO, signal-to-cutoff ratio; SD, standard deviation.

b

CV%, coefficient of variation.