Table 2. Final pullout strength, number of completed cycles at system failure and absolute numbers of anchor pullout as failure mechanism.
Final pullout strength (N) | Completed number of cycles at pullout | Anchor pullout | |
---|---|---|---|
Mean conventional anchor (n = 14) | 259 ± 61 | 4093 ± 1142 | 13/14 |
Mean all-suture anchor (n = 14) | 247 ± 78 | 3791 ± 1349 | 14/14 |
p-value | 0.65 | 0.52 |