Table 3:

Summary of intra-day and inter-day accuracy and precision of LC-MS/MS assay for the determination of valacyclovir (VACV) and acyclovir (ACV) in mouse and human plasma. CV: coefficient of variation.

Analyte and Matrix Concentrations (nM) Intra-day (n=5)
Inter-day (n=5)
Accuracy (%) Precision (CV%) Accuracy (%) Precision (CV%)
ACV in mouse plasma 10; 200; 2000 99.4%; 97.3%; 101.9% 5.5%; 8.0%; 7.6% 98.9%; 96.5%; 104.1% 8.9%; 5.9%; 6.5%
ACV in human plasma 10; 200; 2000 108.1%; 94.8%; 105.8% 4.2%; 2.6%; 5.0% 107.8%; 96.0%; 104.0% 3.8%; 3.0%; 5.5%
VACV in mouse plasma 10; 200; 2000 102.9%; 106.8%; 100.6% 5.2%; 3.7%; 2.1% 105.4%; 104.4%; 102.2% 5.4%; 4.1%; 3.7%
VACV in human plasma 10; 200; 2000 104.1%; 102.4%; 101.6% 5.9%; 2.0%; 4.2% 108.7%; 101.1%; 102.1% 6.4%; 2.8%; 4.5%