Skip to main content
. 2019 Nov 19;2019:5674673. doi: 10.1155/2019/5674673

Table 7.

Final specifications of the designed system.

WS/parameter Specifications
General (i) Rechargeable battery
(ii) Compatible with most android devices
(iii) Small size, portable, and easy to use
(iv) Comfortable for adult and old patients
(v) Multiple parameters: ECG, HR, SpO2, temperature, and RR
(vi) The ability of recording a video and data text of signals
(vii) Sending autoalarm to the doctor/centers in different tools

ECG and HR (i) ECG: single lead without the need for a third right-leg drive (DRL) electrode
(ii) Heart rate detected by R to R distance
(iii) Frequency range: 15.625 mHz up to 256 Hz
(iv) ECG calibration: ±0.25 mV
(v) Arrhythmia analysis: yes

SpO2 (i) Display: waveforms and digits
(ii) Real-time display of PPG (photoplethysmogram)
(iii) Measurement range: 1–100%
(iv) Resolution: 1%
(v) Accuracy: 2% (80–100%)
(vi) Pulse rate range: 20–300 bpm

RR (respiratory rate) (i) Method: air flow temperature
(ii) Measurement range: 5∼50 rpm
(iii) Accuracy: ±1 bpm
(iv) Resolution: 2 bpm

Temperature (i) Measurement range: 0∼50°C
(ii) 0.1°C accuracy (37°C to 39°C)
(iii) Resolution: 0.1°C