Table 3.

The mean temperature gradient at abutment level (P2 thermocouple). The temperature rose significantly with an increased laser fluence. (p < 0.0003). The temperature rise by the critical 10 °C (10.67 + 1.14 °C) was noted only for the ball type fixture (BTF) at 200 mJ/30 Hz. The temperature increase was significantly lower for the scaler (G4) in contrast to Er:YAG laser at FTF. (p < 0.0002).

Study Groups Thermocouple P2∆Ta (°C) IA (A)
(Mean ± SD)
Thermocouple P2∆Ta (°C) FTF (A)
(Mean ± SD)
Thermocouple P2∆Ta(°C) BTF (A)
(Mean ± SD)
P Value
Group 1 0.95 ± 0.55 * 3.47 ± 0.27 *+ 2.40 ± 0.35 + IA vs. FTF vs. BTF p < 0.05
Group 2 2.55 ± 0.58 5.50 ± 0.89 *+ 3.12 ± 0.74 + FTF vs. IA, BTF p < 0.0002
FTF vs. BTF p < 0.0002
Group 3 5.86 ± 0.46 * 7.62 ± 0.74 *+ 10.67 ± 1.14 *+ IA vs. FTF vs. BTF p < 0.05
Group 4 2.00 ± 0.89 2.13 ± 0.23 2.97 ± 0.27 + BTF vs. IA, FTF p < 0.003
FTF vs. BTF p < 0.003
IA vs. FTF p > 0.05
P value G1 vs. G2, G3, G4 p < 0.002
G2 vs. G3 p < 0.0002
G3 vs. G4 p < 0.0002
G2 vs. G4 p > 0.05
G1 vs. G2 vs. G3 vs. G4 p < 0.0002 G3 vs. G2, G4 p < 0.0002
G1 vs. G3 p < 0.0002
G1 vs. G2, G4 p > 0.05
G2 vs. G4 p > 0.05

* Indicates significant differences between laser and control group G4 (ultrasonic scaler); + Indicates significant differences between the two-piece implant (IA) and one-piece implants (FTF, BTF).