Table 3.

OF relative to the MLC_square test

Field size (cm2) Measured OF PRIMO OF (diff.) Acuros OF (diff.)
2 × 2 0.852 0.857 (0.5%) 0.862 (1.2%)
3 × 3 0.902 0.905 (0.3%) 0.907 (0.6%)
5 × 5 0.946 0.944 (−0.3%) 0.947 (0.0%)