. 2019 Aug 7;30(1):383–393. doi: 10.1007/s00330-019-06380-9

Table 2.

Sequence parameters for R2*

Sequence type Gradient-echo sequence, breath-hold
Coil At least at 3 T or in case of high overload use the body coil (coil autoselection should be switched off)
Plane/orientation Axial
Field-of-view 38–40 cm
TR ~ 120 ms
TE initial < 1 ms (as low as possible)
delta-TE 0.25–1.4 ms
Number of echoes 12–20
FA 20°
Options No fat saturation is advised (depending on the calibration formula)
Slice thickness 7–10 mm
Number of slices The maximum allowed, from the spleen to the pancreas