. 2019 Nov 2;11(11):1709. doi: 10.3390/cancers11111709

Table 3.

Temperature increases at 600 sec for eccentric heating target measured with thermistor probes and computed Magnetic Resonance Thermometry (MRT).

Institution Temp. Measurem. Mean Temperature Increase
[ΔT (°C)] at Locations 1 to 4
Mean Temp. Increase ± SD (°C) d(ΔT Location 4- ΔT Location 1) (°C)
ΔT Location 1 ΔT Location 2 ΔT Location 3 ΔT Location 4
1 Thermistor probe 2.6 3.2 4.3 5.9 4.0 ± 1.2 3.3
MRT 2.0 2.7 4.1 5.3 3.5 ± 1.3 3.3
2 Thermistor probe 4.5 4.7 6.1 7.2 5.6 ± 1.1 2.7
MRT 3.8 4.3 6.0 7.0 5.3 ± 1.3 3.2
3 Thermistor probe 2.5 3.0 3.7 4.6 3.5 ± 0.8 2.1
MRT 2.3 2.5 3.9 4.4 3.3 ± 0.9 2.0
4 Thermistor probe 2.7 2.9 3.0 3.6 3.0 ± 0.3 0.9
MRT 2.6 2.8 3.2 3.7 3.0 ± 0.4 1.1
5 Thermistor probe 1.3 1.7 2.8 4.3 2.5 ± 1.2 3.0
MRT 0.6 0.9 3.2 4.6 2.3 ± 1.6 3.9