Published in final edited form as: Med Phys. 2019 Oct 20;46(12):5572–5592. doi: 10.1002/mp.13801

Table A2.

Estimation of the source blur h2 for Hologic Selenia Dimensions system. We used the maximum current of the x-ray tube in the vendor’s user guide2 as the current for each thickness of the breast, ignoring the possible dependence of the current on kV setting. The Hologic system bins 2 × 2 pixels during the image reconstruction. Therefore, the pixel size is 0.14 mm for this system.

Breast Thickness (mm) Kilovoltage (kV) Total Current-Time Product (mAs) Current (mA) Total Exposure Time (s) Exposure Time per PV (s) Source Motion of one PV (mm) Source Blur h2 (mm)
20 26 32 200 0.160 0.011 0.5 0.8
40 29 43 200 0.215 0.014 0.7 1.0
60 33 60 200 0.300 0.020 1.0 1.3
80 38 81 200 0.405 0.027 1.3 1.6