Skip to main content
. 2019 Nov 20;18:e00148. doi: 10.1016/j.plabm.2019.e00148

Table 2.

Analytes with AMR exceptions.

Assay Units Manufacter’s Upper Limit Claim Observed Upper Limit Recovery
ALP U/L 1500 1600 107%
Lactate mmol/L 10 16.5 165%
Prealbumin mg/dL 80 60 75%
U-Glucose mg/dL 700 970 138%
CSF-Glucose mg/dL 800 1000 125%

Analytes with deviations ​> ​±20% above or below the manufacturer’s claimed upper limit of the analytical measurement range. Values recovered from Verichem linearity material for ALP, Lactate, U-Glu and CSF-Glu. Beckman Coulter control material was used for Prealbumin.