Table 4.
Patient size | Preset Family (Scan FOV, focal spot, coverage) | Rot time (s) | Pitch | Min CTDIvol (mGy) | Max CTDIvol (mGy) |
---|---|---|---|---|---|
Small | Large Body, Large, 80 mm | 0.5 | 0.992 | 4.7 | 12.2 |
Medium | Large Body, Large, 80 mm | 0.6 | 0.508 | 10.9 | 28.4 |
Large | Large Body, Large, 80 mm | 1 | 0.508 | 17.2 | 45.5 |
Extra large | Large Body, XLarge, 80 mm | 1 | 0.508 | 49.7 | 53.7 |
This profile is designed for abdomen and pelvis protocols, where patient motion is not an issue but there is concern about artifacts from helical pitches larger than 1.