Skip to main content
. 2019 Nov 5;20(12):169–179. doi: 10.1002/acm2.12754

Table A5.

Medium Body SFOV, 40 mm coverage.

Patient size Preset Family (Scan FOV, focal spot, coverage) Rot time (s) Pitch Min CTDIvol (mGy) Max CTDIvol (mGy)
Small Medium Body, Large, 40 mm 0.5 1.375 5.5 9.9
Medium Medium Body, Large, 40 mm 0.8 1.375 8.4 15.1
Large Medium Body, Large, 40 mm 0.8 0.984 11.8 21.1
Extra large Medium Body, Large, 40 mm 1 0.516 27.6 49.6