Skip to main content
. 2019 Nov 5;20(12):169–179. doi: 10.1002/acm2.12754

Table A7.

Medium Body SFOV, 40 mm coverage, rot time ≤ 0.6 s.

Patient size Preset Family (Scan FOV, focal spot, coverage) Rot time (s) Pitch Min CTDIvol (mGy) Max CTDIvol (mGy)
Small Medium Body, Large, 40 mm 0.5 1.375 5.5 9.9
Medium Medium Body, Large, 40 mm 0.6 0.984 9.2 16.3
Large Medium Body, Large, 40 mm 0.5 0.516 14.8 26.3
Extra large Medium Body, Large, 40 mm 0.6 0.516 17.5 31.1