Skip to main content
. 2019 Nov 5;20(12):169–179. doi: 10.1002/acm2.12754

Table A8.

Medium Body SFOV, 40 mm coverage, rot time ≥ 0.8 s.

Patient size Preset Family (Scan FOV, focal spot, coverage) Rot time (s) Pitch Min CTDIvol (mGy) Max CTDIvol (mGy)
Small Medium Body, Large, 40 mm 0.8 1.375 8.4 15.1
Medium Medium Body, Large, 40 mm 0.8 0.984 11.8 21.1
Large Medium Body, Large, 40 mm 1 0.984 14.5 26.0
Extra large Medium Body, Large, 40 mm 1 0.516 27.6 49.6