Skip to main content
. 2019 Nov 29;20(12):119–126. doi: 10.1002/acm2.12776

Table 2.

Results of the film dosimetry (relative to water equivalent material alone).

Sample 6 MV 10 MV
Film 1 Film 2 Film 3 Film 1 Film 2 Film 3
FRC‐BG (3 mm) 0% 0% 0% 0% 0% 0%
FRC (3 mm) 4% 1% 0% 6% 0% 0%
PEEK (3.8 mm) 10% 0% −1% 7% 1% 1%
Ti (3 mm) 18% −15% −9% 17% −5% −4%

FRC‐BG, glass fiber reinforced composite with bioactive glass; FRC, glass fiber reinforced composite; PEEK, polyether ether ketone; Ti, titanium.