Experimental limits of detection (LODs) for all organic explosives investigated in this work (Reported LOD values are estimates based on a dilution series and correspond to the lowest concentration for which the main adduct ions for the explosive compound could still be clearly discerned from the background noise)

Explosive class Explosive standard Experimental LOD (ng/mL)
Nitrate esters ETN 1
PETN 10
NG 10
Nitroamines HMX 1
RDX 1
Nitroaromatics TNT 10
Tetryl 10
Peroxide-based HMTD 30,000
TATP 15,000