Table 3.

Comparison of key protocol settings contributing to dose differences between maximum and minimum DLP75kg systems

Examination Clinical purpose of CT Max/min DLP75kg Max or min DLP75kg protocol Scan length (cm) Summary of main protocol differences contributing to variation in DLP75kg
PET/CT brain Localisation/characterisation 16.8 Max 18 Exceptionally high effective mAs for localisation/characterisation (images originally intended for diagnostic use)
Min 23
PET/CT and SPECT/CT cardiac AC only 19.6 Max 29 High effective mAs for AC only; long scan length
Min 16 Attempted to optimise for AC only with low effective mAs (low mA, high pitch) and low kV; use of IR
SPECT/CT lung AC only 27.4 Max 32 High kV; high effective mAs for AC only
Min 36 Attempted to optimise for AC only with low effective mAs (low mA, high pitch); use of IR
Localisation/characterisation 4.2 Max 32 Higher kV and effective mAs than minimum DLP75kg protocol
Min 33 Low kV and effective mAs
SPECT/CT bone Localisation/characterisation 10.0 Max 106 Three times greater scan length, higher kV and effective mAs than minimum DLP75kg protocol
Min 37 Use of IR
SPECT/CT parathyroid Localisation/characterisation 7.8 Max 28 Much higher effective mAs (but lower kV) than minimum DLP75kg protocol
Min 24