Table 1.
Technique | Physical parameter | Sensing parameter | Applications | References | |
---|---|---|---|---|---|
Optical | LSR | G: viscoelastic modulus τ: time constant of laser speckle intensity fluctuations |
Significant correlation between τ and |G*| (r = 0.79, p < 0.0001) | PT and APTT | Tripathi et al., 2014 |
OCE,OCT | μ: shear modulus | Viscoelastic peroperties | Clot shear modulus | Xu et al., 2016 | |
Fluorescent imaging | Fluorescent intensity | Heparin and tinzaparin from 0 to 0.8 U/ml | APTT, PT Thrombodynamics analysis (TD) |
Dudek et al., 2010; Yun et al., 2019 | |
SPR | ns: refractive index | 7 to 8 ng/mm2 cell adhesion detection | PT | Hansson et al., 1999, 2007 | |
Electro-mechanical | Acoustic wave devices (QCM and SAW) | 1. Frequency shift as a result of mass change (μg) in liquid Δm 2. Dissipation factor in liquid ΔD |
Factor VIII: 0.0185 mg/L Fibrinogen: 1 g/Lf |
PT, APTT, and INR | Rodahl et al., 1995 Yao et al., 2013 Hussain, 2015 |
Microfluidic viscometers | Pressure (Pa) and mechanical displacements | ~0.4–12 p.s.i. Heparin detection: 0.25 IU/mL |
PT, APTT, ACT, and INR | Jain et al., 2016a; Hegener et al., 2017; Yeom et al., 2017; Judith et al., 2018 Kang et al., 2019 |
|
MEMS (Magneto elastic cantilevers) | Viscosity | Blood viscosity: 0.08 cp | PT, APTT, and INR | Cakmak et al., 2015 | |
Ultrasound transducers | Sound velocity (m/s) Displacement (μm) |
Sensitivity of 3 cm/s | Viscoelastic test–derived coagulation parameters | Voleišis et al., 2002; Huang C.-C. et al., 2011; McLoughlin et al., 2016 | |
Photoacoustic detection | Photoacoustic signal (PA intensity) μa, μeff, Γ | 0.28 IU/mL of heparin detection | APTT | Wang et al., 2016 | |
Electrochemical (Amperometric, impedance analysis) |
Conductance, electrical current (amplitude/phase) [Ohm] | 0.33 fM (thrombin detection) | PT, APTT, and INR | Lei et al., 2013; Liu et al., 2018 |