Table 1.

Summary of the blood coagulation measurement techniques and physical parameters in each technique in research methodologies.

Technique Physical parameter Sensing parameter Applications References
Optical LSR G: viscoelastic modulus
τ: time constant of laser speckle intensity fluctuations
Significant correlation between τ and |G*| (r = 0.79, p < 0.0001) PT and APTT Tripathi et al., 2014
OCE,OCT μ: shear modulus Viscoelastic peroperties Clot shear modulus Xu et al., 2016
Fluorescent imaging Fluorescent intensity Heparin and tinzaparin from 0 to 0.8 U/ml APTT, PT
Thrombodynamics analysis (TD)
Dudek et al., 2010; Yun et al., 2019
SPR ns: refractive index 7 to 8 ng/mm2 cell adhesion detection PT Hansson et al., 1999, 2007
Electro-mechanical Acoustic wave devices (QCM and SAW) 1. Frequency shift as a result of mass change (μg) in liquid Δm
2. Dissipation factor in liquid ΔD
Factor VIII: 0.0185 mg/L
Fibrinogen: 1 g/Lf
PT, APTT, and INR Rodahl et al., 1995
Yao et al., 2013
Hussain, 2015
Microfluidic viscometers Pressure (Pa) and mechanical displacements ~0.4–12 p.s.i.
Heparin detection: 0.25 IU/mL
PT, APTT, ACT, and INR Jain et al., 2016a; Hegener et al., 2017; Yeom et al., 2017; Judith et al., 2018
Kang et al., 2019
MEMS (Magneto elastic cantilevers) Viscosity Blood viscosity: 0.08 cp PT, APTT, and INR Cakmak et al., 2015
Ultrasound transducers Sound velocity (m/s)
Displacement (μm)
Sensitivity of 3 cm/s Viscoelastic test–derived coagulation parameters Voleišis et al., 2002; Huang C.-C. et al., 2011; McLoughlin et al., 2016
Photoacoustic detection Photoacoustic signal (PA intensity) μa, μeff, Γ 0.28 IU/mL of heparin detection APTT Wang et al., 2016
Electrochemical (Amperometric,
impedance analysis)
Conductance, electrical current (amplitude/phase) [Ohm] 0.33 fM (thrombin detection) PT, APTT, and INR Lei et al., 2013; Liu et al., 2018