Table III.
τi, ns | Fi | ||||
---|---|---|---|---|---|
sample | expecteda | found | expectedb | found | χR2 |
Per + 9-AAc (50/50) | 8.95 | 1.0 | 38.0 | ||
5.07c | 5.26d | 0.42 | 0.38 | ||
14.18 | 14.89 | 0.58 | 0.62 | 2.9 | |
Per + 9-AAe (07/93) | 13.31 | 1.0 | 4.0 | ||
5.07 | 5.50 | 0.07 | 0.05 | ||
14.18 | 13.94 | 0.93 | 0.95 | 2.3 | |
p-QP + PPOc,g (50/50) | 1.12f | 1.0 | 7.7 | ||
0.80 | 0.81 | 0.48 | 0.46 | ||
1.42 | 1.53 | 0.52 | 0.54 | 2.5 |
From the usual FD measurements.
From the steady-state spectra normalized to 1.0.
The FD measurements of Per + 9-AA were performed using a Corning 3–74 emission filter, and of p-QP + PPO using a WG filter. [9-AA] = [Per] = 5 × 10−6 M.
Measured at 12 modulation frequencies.
[9-AA] = 1 × 10−5 M, [Per] = 1 × 10−6 M.
Measured at 16 modulation frequencies.
[p-QP] = [PPO] = 3 × 10−6 M.