Table 3. Table comparing the results of Chagas Stat Pack (CSP) and Chagas Detect Plus (CDP) individually or taking into consideration their combined use (RDTs) versus the "gold-standard" based on ELISAs.

ELISAs* CSP CDP RDTs#
False positive 0 10 49 15
True positive 304 297 299 297
False negative 0 7 5 7
True negative 381 371 332 366
Total 685 685 685 685
% Agreement - 97.5 92.1 96.8
Kappa (CI95%) - 0.95 (0.93–0.97) 0.84 (0.80–0.88) 0.94 (0.91–0.96)
Sensitivity - 97.7 98.4 97.7
Specificity - 97.4 87.1 96.1
PPV - 96.7 85.9 95.2
NPV - 98.1 98.5 98.1
DE - 97.5 92.1 96.8

Note that CSP, CDP and RDTs results of “% of agreement” (and kappa value), “sensitivity”, “specificity”, “PPV”, “NPV” and “DE” shown in the table correspond to their comparison to the outcome of using the ELISAs.

*, results obtained with the standard algorithm based on the agreement of two conventional serological tests (ELISAs in this case);

#, results obtained with the combined use of up to three RDTs, as described.