Table II.
Transducer | Biconcave 1–3 piezoelectric composite [24] | Valpey Fisher IS0504HR | Panametrics ZF3001 | Olympus P150-2-R2.00 |
---|---|---|---|---|
Diameter (cm) | 4 | 1.27 | 0.635 | 0.635 |
Center Frequency (MHz) | 2 | 5 | 30 | 50 |
Maximum Sweep Frequency (MHz) | 20 | 20 | 40 | 40 |
Sweep Time (s) | 0.2 | 0.334 | 0.334 | 0.267 |
Sweep Rate S (MHz/s) | 100 | 60 | 120 | 150 |
Hydrophone Depth (z) (cm) | 15 | 30 | 15 | 10 |
Delay Time tD=z/c (μs) | 101 | 203 | 101 | 68 |
TDS Frequency Δf = StD (kHz) | 10.1 | 12.1 | 12.1 | 10.1 |
Filter Bandwidth (BW) (kHz) | 0.67 | 0.4 | 0.8 | 1 |
Frequency Resolution = S/BW (MHz) | 0.15 | 0.15 | 0.15 | 0.15 |
Time Resolution = BW / S (μs) | 6.7 | 6.7 | 6.7 | 6.7 |