Table II.

Transducer properties and TDS parameters. The source frequency was swept linearly from 0 to the maximum sweep frequency. See [11] for relationships among TDS parameters.

Transducer Biconcave 1–3 piezoelectric composite [24] Valpey Fisher IS0504HR Panametrics ZF3001 Olympus P150-2-R2.00
Diameter (cm) 4 1.27 0.635 0.635
Center Frequency (MHz) 2 5 30 50
Maximum Sweep Frequency (MHz) 20 20 40 40
Sweep Time (s) 0.2 0.334 0.334 0.267
Sweep Rate S (MHz/s) 100 60 120 150
Hydrophone Depth (z) (cm) 15 30 15 10
Delay Time tD=z/c (μs) 101 203 101 68
TDS Frequency Δf = StD (kHz) 10.1 12.1 12.1 10.1
Filter Bandwidth (BW) (kHz) 0.67 0.4 0.8 1
Frequency Resolution = S/BW (MHz) 0.15 0.15 0.15 0.15
Time Resolution = BW / S (μs) 6.7 6.7 6.7 6.7