Table 1.
Scanning Event | Device | Indication | Scan Type | Post‐MRI ILR Status |
---|---|---|---|---|
1 | Reveal Plus 9526 | Head trauma | GE 1.5 T Cranial | Normal function Artifact: none |
2 | Reveal Plus 9526 | Shoulder trauma | GE 1.5 T Shoulder | Normal function Artifact: Wide and narrow complex tachycardias |
3 | Reveal Plus 9526 | Suspected brain mass | GE 1.5 T Cranial | Normal function Artifact: none |
4 | Reveal Plus 9526 | Confusion | GE 1.5 T Lumbar‐spine | Normal function Artifact: SVT |
5 | Reveal Plus 9526 | Headache | GE 1.5 T Cranial | Normal function Artifact: CHB, SVT |
6 | Reveal Plus 9526 | Evaluation of multiple sclerosis | GE 1.5 T Cranial | Normal function Artifact: SVT |
7 | Reveal Plus 9526 | Suspected brain mass | GE 1.5 T Cranial | Normal function Artifact: none |
8 | Reveal Plus 9526 | Weakness | GE 1.5 T Cranial | Normal function Artifact: SVT |
9 | Reveal Plus 9526 | Left knee pain | GE 1.5 T Knee | Normal function Artifact: none |
10 | Reveal Plus 9526 | Left leg pain | GE 1.5 T Lumbar spine | Normal function Artifact: asystole, SVT |
11 | Reveal Plus 9526 | Suspected brain mass | GE 1.5 T Cranial | Normal function Artifact: SVT |
MRI = magnetic resonance imaging; ILR = insertable loop recorder; T = Tesla; SVT = supraventricular tachycardia; CHB = complete heart block; GE = General Electric, Corp.