Table 3.

Typical components that might be used to construct a fluorescence detection pathway. Example specifications are designed for fluorescein-tagged compounds with an excitation wavelength near 488 nm and emission wavelength near 525 nm.

Component name Selection considerations Example specifications

Excitation Laser Single wavelength, low power, continuous wave Wavelength-488 nm
Power-20 mW

Single mode fiber optic Consistent with laser wavelength Operating wavelength-405–532 nm

Dichroic mirror Reflect excitation light and pass emission light Transmission band-520–800 nm
Reflection band-380–490 nm
Cut-on wavelength-505 nm

Microscope objective High NA and compatible with filter cube Magnification-40x
NA-0.65

Multi-mode fiber optic Consistent with emission wavelength and large core diameter Core diameter-25 ± 3 μm
Wavelength range-400–550 nm

Photo multiplier tube (PMT) Large dynamic range and high sensitivity Wavelength range-300–720 nm
Sensitivity-176 mA/W

Fiber optic couplers Adjustable in three dimensions and low hysteresis

Collimator Couples directly to fiber optic

Filter cube Compatible with dichroic, couplers, objective, and collimator