Table 4.

Fused silica capillary parameters.

Parameter Available Options Recommendation Rationale
Capillary OD 90 – >3500 μm 360 μm Dissipates heat well (reduces Joule heating); less prone to breaking
Capillary ID ≤1 – >2000 μm 30 – 50 μm Minimizes physical blockages; high resistance yields low currents
Total Length User preference ≤ 40 cm Longer yields higher separation efficiencies and resolution
Effective Length ≤ Total length < 20 cm Shorter yields higher throughput; limited by instrument design
Detection Window Size User preference ≤ 0.5 cm Enables facile z-positioning of window; limits window breakage