Table 3.

Contrast-noise ratio (CNR) for volumetric imaging and steak artefact index (SI) on X-ray volumetric imaging for different fiducial markers

Volumetric imaging Visibility (CNR) Artefact (SI)
RP CIVCO GA GA (B) PM RP CIVCO GA GA(B) PM
CT (slice thickness (mm))
 0.625 1.46 1.49 1.53 1.54 1.47 67.78 110.9 61.04 67.51 25.57
 1.25 1.61 1.46 1.44 1.22 1.52 86.92 52.81 65.20 59.64 66.24
 2.5 1.27 1.23 1.42 1.34 1.52 53.35 47.19 36.88 46.15 11.63
CBCT (protocol)
 P Varian 1.17 0.96 0.99 0.92 0.79 33.16 22.03 33.81 13.86 19.55
 P Medium 0.94 0.94 0.64 1.07 0.90 26.38 42.57 15.58 19.91 9.82
 P Obese 1.00 0.84 0.85 1.16 1.27 66.08 58.28 47.48 75.81 36.88
MRI FM size (mm3)
P 13.57 13.57 5.03 5.03 9.42
 T1w 1.17 1.69 1.77 1.64 2.56 6.67 10.00 6.67 26.67 3.33
 T2w (filtered) 1.95 1.87 1.54 1.45 1.88 13.33 20.00 16.67 23.33 13.33
 VIBE 1.58 1.94 1.57 1.47 2.10 53.33 70.0 123.33 136.67 110.00

Note that CNR and SI values are calculated for each box as a contribution of the 3 FMs contained in each box while FM size in MRI corresponds to one FM to facilitate the comparison with their physical dimensions. P indicates the physical size of the markers calculated assuming the markers are perfect cylinders