Table 5.

Accuracy and intra- and interday precision of the standard compounds in MHT.

Compound Spiked amount (mg/mL) Content (mg/mL) Recovery test (%, n = 5) Precision test (n = 5)
Expected Measured Intraday RSDa (%) Interday RSD (%)
1 0.025 0.0929 0.0929 100.2 0.89 0.29
0.05 0.1179 0.1192 102.6 0.58 0.59
0.1 0.1679 0.1708 102.9 0.13 0.25
5 0.025 0.0349 0.0348 99.6 0.83 0.24
0.05 0.0599 0.0606 101.3 0.34 0.40
0.1 0.1099 0.1117 101.8 0.06 0.18
6 0.025 0.0328 0.0322 97.8 0.68 0.39
0.05 0.0578 0.0582 100.7 0.10 0.05
0.1 0.1078 0.1087 101.0 0.04 0.11
7 0.025 0.0261 0.0267 102.3 2.06 0.18
0.05 0.0511 0.0523 102.4 0.12 0.10
0.1 0.1011 0.1029 101.8 0.14 0.07

aRSD: relative standard deviation.