Skip to main content
. 2019 Dec 23;58(1):e00888-19. doi: 10.1128/JCM.00888-19

TABLE 5.

Commutability of EBV assay pairs by linear regression method—Exact reference material

Fragmentation and assay No. of samples for which results agreed/total no. of samples testeda
EliTech Qiagen Altona (7500) Luminex DiaSorin LDT (eBER) LDT
(IR-1)
Altona
(digital)
LDT EBER
(digital)
Nonfragmented
    Qiagen 5/5
    Altona (7500) 5/5 5/5
    Luminex 5/5 5/5 4/5
    DiaSorin 4/4 4/4 4/4 4/4
    LDT (EBER) 4/4 4/4 3/4 4/4 4/4
    LDT (IR-1) 5/5 5/5 0/5 4/5 1/4 4/4
    Altona (digital) 5/5 5/5 5/5 4/5 4/4 3/4 0/5
    LDT (EBER-digital) 4/4 4/4 4/4 4/4 4/4 4/4 4/4 4/4
    LDT (IR-1-digital) 5/5 5/5 2/5 5/5 4/4 4/4 5/5 4/4 0/5
Fragmented
    Qiagen 5/5
    Altona (7500) 5/5 5/5
    Luminex 5/5 5/5 5/5
    DiaSorin 4/4 4/4 4/4 4/4
    LDT (EBER) 5/5 5/5 4/5 5/5 4/4
    LDT (IR-1) 5/5 5/5 1/5 4/5 0/4 4/5
    Altona (digital) 5/5 5/5 5/5 5/5 4/4 2/5 0/5
    LDT (EBER-digital) 3/3 3/3 3/3 3/3 3/3 3/3 3/3 3/3
    LDT (IR-1-digital) 5/5 5/5 2/5 4/5 2/4 4/5 5/5 0/5 3/3
a

Results for commutable reference material are in bold and those for noncommutable reference material are shaded; otherwise, the data are for marginally commutable reference material. Numerators represent the number of commutable tests, while the denominators are the number of total tests run.