. 2019 Dec 18;3(2):e9815. doi: 10.2196/cardio.9815

Table 2.

Rate of transmission, overall and by priority.

Transmission priority All (n=153, 107 patient-years) CRT-Da (n=76, 53 patient-years) ICDb (n=15, 10 patient-years) IPGc (n=61, 61 patient-years) CRT-Pd (n=1, 1 patient-year)
All transmission

Transmissions, n 654 426 50 176 2
Annual rate of transmissions per 100 patient-years (95% CI) 613 (568-662) 802 (729-882) 504 (382-665) 410 (354-476) e
No prioritized event (green transmissions)

Transmissions, n (%) 551 (84.3) 364 (85.4) 40 (80.0) 147 (83.5) 0 (0.0)
Patients with green transmission, n 141 67 14 60 0
Annual rate of transmissions per 100 patient-years (95% CI) 517 (475-561) 648 (585-718) 393 (288-536) 333 (283-391)
Low-priority events (amber transmissions)

Transmissions, n (%) 88 (13.5) 53 (12.4) 9 (18.0) 24 (13.6) 2 (100)
Patients with amber transmission, n 50 28 4 17 1
Annual rate of transmissions per 100 patient-years (95% CI) 82.5 (66.9-102) 94.4 (72.1-124) 88.4 (46.0-170) 54.3 (36.4-81.1) 277 (69.2-1106)
High priority events (red transmissions)

Transmissions, n (%) 15 (2.2) 9 (2.2) 1 (2.0) 5 (1.7) 0 (0.0)
Patients with red transmission, n 9 5 1 3 0
Annual rate of transmissions per 100 patient-years (95% CI) 14.1 (8.5-23.3) 16.0 (8.3-30.8) 9.8 (1.4-69.7) 11.3 (4.7-27.2)

aCRT-D: cardiac resynchronization therapy defibrillator.

bICD: single- or dual-chamber implantable cardioverter defibrillator.

cIPG: single- or dual-chamber pacemaker.

dCRT-P: cardiac resynchronization therapy pacemaker.

eNot applicable.