Engineering metrics of the robotic and OCT systems

The IRISS Robotic System 17,18
Positional precision* 27±2 µm
Positional accuracy 205±3 µm
Robot-to-eye alignment time < 1 min
Mounted tool 8172 UltraFLOW straight-tip irrigation/aspiration handpiece with side aspiration port (Alcon, Fort Worth, TX)
The OCT System
Detection scheme Spectral Domain
Model Telesto II 1060LR with objective lens LSM04BB (ThorLabs, Newton, NJ)
Central wavelength 1060 nm
Volume scan dimensions 10×10×9.4 mm
Volume scan acquisition time 33.2 s
Axial resolution (in air) 9.18 µm
B-scan acquisition and display rate 4.65 Hz
*

“Positional precision” refers to the ability to repeatably touch the same point

“Positional accuracy” refers to the ability to exactly touch a specified point