Skip to main content
. 2019 Dec 23;2019:2018598. doi: 10.1155/2019/2018598

Table 1.

Analytic parameters of myocardial enzyme activity on the VITROS 5600 Integrated System.

Enzyme SI unit Bias of accuracy (%) Precision Analytical measurement range (U/L) Clinical reportable range (U/L) Accreditation criteria
Low level (%) High level (%)
Within-day Between-day Within-day Between-day Bias (%) TE (%)
AST U/L 4.86 1.40 1.42 1.40 1.52 9.0–653.0 9.0–6530.0 5.00 15.00
LDH U/L 1.91 2.30 2.33 1.85 1.90 294.0–1473.0 294.0–14730.0 4.00 11.00
CK U/L 4.85 2.15 4.20 1.79 3.27 21.5–1566.0 21.5–1566.0 5.50 15.50
CKMB U/L 5.51 5.18 5.20 4.30 4.31 2.7–300.0 2.7–300.0 7.00 22.00

TE refers to total error; Bias and TE criteria are from Analytical quality specification for routine analytes in clinical biochemistry published by the Ministry of Health of the People's Republic of China; due to lack of criteria for CKMB, bias and TE were recommended by our own clinical laboratory. Coefficient of variation (CV) for within-day precision should be < 1/4TE, and CV for between-day precision should be < 1/3TE according to Guidance on the Application of Accreditation Criteria for the Medical Laboratory Quality and Competence in the Field of Clinical Chemistry published by the China National Accreditation Service for Conformity Assessment.