Table 3.

The National Electrical Manufactures Association Image quality test result, i.e., contrast recovery, background variability, and percentage relative lung error for the images reconstructed by VPHD and Q. Clear reconstruction technique

Image quality Hot sphere
Cold sphere
10 mm 13 mm 17 mm 22 mm 28 mm 37 mm
The percentage contrast for each hot sphere
 Prescribed limits (>30) (>40) (>50) (>60) (>60) (>60)
 VPHD 33.7 43.1 59.9 67.7 71.5 74.2
 Q. Clear 41.9 59.4 69.7 85 82.1 88.9
The percentage background variability
 Prescribed limits (<12) (<10) (<9) (<7) (<6) (<5)
 VPHD 5.7 4.7 3.7 2.8 2.2 1.8
 Q. Clear 4.8 4.1 3.2 2.4 1.8 1.3
Accuracy of attenuation and scatter corrections
 Prescribed limits (<5.5)
 VPHD 3.74
 Q. Clear
Percentage relative error for each slice in lung slice
 Prescribed limits (19)
 VPHD 16.8
 Q. Clear 10.1
Radioactivity concentration in KBq/mL 5.3 KBq/mL