Table 3.

Pearson correlation coefficients for type III and type IV PM devices

  AHI (PM) AHI (PSG) AI (PM) AI (PSG) HI (PM) HI (PSG)
Correlation between type III PM devices and PSG testing
AHI (PM) 1.00 0.92 0.93 0.94 0.22 – 0.32
AHI (PSG) 0.92 1.00 0.85 0.92 0.2 – 0.08
AI (PM) 0.93 0.85 1.00 0.95 – 0.15 – 0.48
AI (PSG) 0.94 0.92 0.95 1.00 0.01 – 0.46
HI (PM) 0.22 0.2 – 0.15 0.01 1.00 0.43
HI (PSG) – 0.32 – 0.08 – 0.48 – 0.46 0.43 1.00
Correlation between type IV PM devices and PSG testing
AHI (PM) 1.00 0.69 0.97 0.63 – 0.07 0.02
AHI (PSG) 0.69 1.00 0.73 0.9 – 0.31 0.05
AI (PM) 0.97 0.73 1.00 0.68 – 0.3 – 0.02
AI (PSG) 0.63 0.9 0.68 1.00 – 0.35 – 0.39
HI (PM) – 0.07 – 0.31 – 0.3 – 0.35 1.00 0.14
HI (PSG) 0.02 0.05 – 0.02 – 0.39 0.14 1.00

AHI (PM), AI (PM), and HI (PM) manifested the result of type III or type IV PM devices; AHI (PSG), AI (PSG), and HI (PSG) manifested the result of PSG. AHI, apnea-hypopnea index; AI, apnea index; HI, hypopnea index; PM, portable monitor; PSG, polysomnography.