Skip to main content
. 2018 Jun 26;3(2):1–10. doi: 10.1159/000489158

Table 4.

AHI, AI, and HI characteristics as a Brand-Altman Plot

  Type III PM device Type IV PM device
AHI (PM) – AHI (PSG)
 Lower limit line – 26.67 (– 32.88, – 20.45) – 38.07 (– 47.57, – 28.57)
 Mean difference – 8.52 (– 12.11, – 4.93) – 6.29 (– 11.78, – 0.81)
 Upper limit line 9.62 (3.40, 15.84) 25.48 (15.98, 34.98)
AI (PM) – AI (PSG)
 Lower limit line – 27.54 (– 34.49, – 20.59) – 32.15 (– 42.59, – 21.71)
 Mean difference – 7.26 (– 11.27, – 3.25) 2.76 (– 3.27, 8.79)
 Upper limit line 13.02 (6.07, 19.97) 37.67 (27.23, 48.10)
HI (PM) – HI (PSG)
 Lower limit line – 18.19 (– 24.00, – 12.39) – 30.40 (– 36.78, – 24.02)
 Mean difference – 1.26 (– 4.61, 2.09) – 9.06 (– 12.74, – 5.37)
 Upper limit line 15.67 (9.87, 21.47) 12.29 (5.91, 18.67)

Values as given as means (confidence interval). AHI (PM), AI (PM), and HI (PM) show the results from type III or type IV PM devices; AHI (PSG), AI (PSG), and HI (PSG) show the results of PSG testing. AHI, apnea-hypopnea index; AI, apnea index; HI, hypopnea index; PM, portable monitor; PSG, polysomnography.