Percentage of time the devices displayed a value during the evaluated physiological noise conditions

Physiological noises RRoxi (mean ± SD) RRa (mean ± SD) p-value
Total (%) 82.4 ± 19.2 58.0 ± 14.8 < 0.001**
Groaning (%) 96.6 ± 10.4 81.6 ± 20.7 0.001**
Snoring (%) 84.2 ± 31.1 72.6 ± 32.8 0.075
Talking (%) 79.3 ± 32.7 36.5 ± 24.6 <0.001**
Coughing (%) 56.9 ± 32.3 18.9 ± 25.7 < 0.001**

*p-value < 0.05

**p-value < 0.01