Skip to main content
. 2020 Jan 9;10(1):2. doi: 10.1038/s41408-019-0267-8

Table 2.

Concordance of FLC measurements.

(A) Concordance of κ FLC
κ FLC N Latex FLC κ FLC Sebia FLC κ FLC Sebia FLC
Freelite Normal Abnormal Freelite Normal Abnormal N Latex FLC Normal Abnormal
Normal 70 (37%) 11 (6%) Normal 64 (34%) 12 (7%) Normal 65 (35%) 16 (9%)
Abnormal 6 (3%) 100 (54%) Abnormal 13 (7%) 98 (52%) Abnormal 12 (6%) 94 (50%)
Cohens Kappa coefficient: 0.81 Cohens Kappa coefficient: 0.72 Cohens Kappa coefficient: 0.69
(B) Concordance of λ FLC
λ FLC N Latex FLC λ FLC Sebia FLC λ FLC Sebia FLC
Freelite Normal Abnormal Freelite Normal Abnormal N Latex FLC Normal Abnormal
Normal 50 (27%) 22 (12%) Normal 44 (23%) 28 (15%) Normal 40 (21%) 26 (14%)
Abnormal 16 (8%) 99 (53%) Abnormal 18 (10%) 97 (52%) Abnormal 22 (12%) 99 (53%)
Cohens Kappa coefficient: 0.56 Cohens Kappa coefficient: 0.47 Cohens Kappa coefficient: 0.43
(C) Concordance of κ/λ ratio
κ/λ ratio N Latex FLC κ/λ ratio Sebia FLC κ/λ ratio Sebia FLC
Freelite Normal Abnormal Freelite Normal Abnormal N Latex FLC Normal Abnormal
Normal 42 (22%) 11 (6%) Normal 34 (18%) 19 (10%) Normal 39 (21%) 13 (7%)
Abnormal 10 (5%) 124 (67%) Abnormal 17 (9%) 117 (63%) Abnormal 12 (6%) 123 (66%)
Cohens Kappa coefficient: 0.72 Cohens Kappa coefficient: 0.52 Cohens Kappa coefficient: 0.66

Assay-specific reference ranges were used for the classification of a normal (within reference range) and abnormal (outside reference range) FLC. Determination of κ, λ, and κ/λ ratio concordances were performed using 187 serum samples