Table 1.

Q measurements on candidate 17O coils with various apertures and tuning components. The coil length was fixed at 27 cm, with apertures to emulate a single coil in a 4 (90°) or 8 channel (45°) array. The coil with 90° aperture and 2 distributed capacitors represents the baseline (row 1), while the others show compromised performance primarily due to increased capacitor loss (row 2) or reduced coil-to-sample coupling (row 3). This information motivated the implemented design (row 4), in which the aperture provided sufficient coil-to-sample coupling, while loss due to 1H blocking circuits was considered a necessary penalty to enable 1H body coil excitation.

Aperture Number of
Distributed
capacitors
Unloaded Q

Qun
Loaded Q

Ql
Ratio

Qun/Ql
90° 2 420 190 2.21
90° 4 320 180 1.78
45° 2 285 170 1.68
90° 1 first-order and 2 second-order blocking circuits 320 180 1.78