Skip to main content
. 2019 Dec 3;11(12):704. doi: 10.3390/toxins11120704

Table 7.

Recovery data of the employed analytical method based on solvent (RA) and matrix-matched (RE) calibration curves and matrix effect (SSE).

Analytes Concentration Range
(µg kg−1)
Overall Method Recovery RA (%) * Sample Preparation Recovery RE (%) ** Matrix Effect
SSE (%) ***
LOD/LOQ (µg kg−1)
AOH 2.5–100 90.4 80.4 112.5 0.75/2.5
AME 2.5–100 68.5 83.0 82.5 0.3/0.9
TEN 2.5–100 90.4 74.1 122.0 0.5/1.5

AOH—alternariol; AME—alternariol monomethyl ether; TEN—tentoxin; * Calculated by (slope of spiked sample-prepared curve/slope of solvent calibration curve); ** Calculated by (slope of spiked sample-prepared curve/slope of matrix-matched calibration curve); *** Calculated by (slope of matrix-matched calibration curve/slope of solvent calibration curve).