. 2019 Dec 2;6(6):220–225. doi: 10.1049/htl.2019.0074

Table 2.

Measurement of the diameter of six cylindrical inserts in an ultrasound image of a calibration phantom Gammex 403GS LE. All diameters were measured both horizontally (H) and vertically (V)

Phantom, mm QLAB, mm Tomtec, mm VR, mm
Size H/V Mean SD Mean SD Mean SD
4.0 V 3.4 0.508 2.8 0.837 4.4 0.449
4.0 H 2.9 0.397 3.2 0.447 3.8 0.778
6.0 V 5.3 0.219 4.2 0.837 5.9 0.770
6.0 H 4.9 0.241 5.0 0.000 6.3 1.113
10.0 V 10.3 0.652 9.0 0.707 10.4 0.920
10.0 H 9.6 0.205 10.4 0.548 11.7 1.882

Mean result closest to the reference and the smallest standard deviations are shown in bold.