Transverse Spatial Resolution of PennPET Explorer Whole-Body Imager

Parameter Radius (cm) Rings (n) Algorithm Radial (mm) Tangential (mm) Axial (mm)
FWHM 1 1 Analytic 4.2 ± 0.3 3.9 ± 0.4 4.1 ± 0.2
1 1 Iterative 3.9 ± 0.3 3.8 ± 0.3 3.6 ± 0.2
1 3 Iterative 3.9 ± 0.2 3.9 ± 0.3 3.9 ± 0.3
10 3 Iterative 4.2 ± 0.2 3.9 ± 0.2 3.9 ± 0.3
20 3 Iterative 5.6 ± 0.2 3.9 ± 0.4 3.7 ± 0.3
FWTM 1 1 Analytic 8.5 ± 0.8 8.4 ± 0.9 7.9 ± 0.2
1 1 Iterative 7.2 ± 0.4 7.1 ± 0.2 6.8 ± 0.2
1 3 Iterative 7.4 ± 0.6 7.3 ± 0.2 7.8 ± 1.2
10 3 Iterative 8.1 ± 0.2 7.2 ± 0.2 7.6 ± 0.8
20 3 Iterative 10.4 ± 0.3 7.2 ± 0.2 7.3 ± 0.7

FWHM = full width at half maximum; FWTM = full width at tenth maximum.