Table 1.
Alert Type | Qualifying Event Category | IMD Function | EXD Function | Data Storage |
---|---|---|---|---|
High Priority Alarms | ||||
Emergency alerts (Patient calls 911 for prompt arrival at a medical facility | 3 successive 10-second electrogram segments with each segment at least six out of eight beats with ST shifts that exceed the detection threshold while the heart rate is within normal range | Vibrates | Beeps in a synchronized 3-2-3-2 pattern, and flashes a red light | 24 hrs before alarm through 8 hrs after alarm |
Low Priority Alarms | ||||
See Doctor alerts | (a) Abnormal heart rates (Tachycardia and bradycardia) (b) Persistent irregular rhythms (c) Loss of signal (Lead detachment/low battery/general malfunction) |
Vibrates | Beeps once after each 7 second interval, and flashes a yellow light | 24 hrs before alarm and at time of alarm |
None | Save data in the See Doctor manner but do not alert the patient | N/A | N/A | 24 hrs before alarm and at time of alarm |
Ignore | Neither save the data nor alert the patient | N/A | N/A | N/A |