Comparison of analytical parameters for different biosensors: Ty/bare SPE, Ty/Biochar/SPE, and Ty/Graphene/SPE biosensors.

Linear Range
(mM)
LOD
(mM)
KMapp
(mM)
Ty/biochar/SPE 0.02–0.50 2.4 × 10−4 0.15
Ty/graphene/SPE 0.02–0.25 1.0 × 10−4 0.09
Ty/bare SPE 0.02–2.0 9.2 × 10−4 0.25