Table 2. Image Quality Scores.
Parameters | C-DWI | SMS2-DWI | SMS3-DWI | P* | P†: C vs. SMS2 | P†: SMS2 vs. SMS3 | P†: C vs. SMS3 |
---|---|---|---|---|---|---|---|
Sharpness | 3.54 ± 0.47 | 3.58 ± 0.40 | 3.02 ± 0.50 | < 0.001 | > 0.999 | < 0.001 | < 0.001 |
Distortion | 2.76 ± 0.51 | 2.78 ± 0.48 | 2.78 ± 0.48 | 0.526 | |||
Artifact | 3.48 ± 0.40 | 3.42 ± 0.37 | 2.84 ± 0.38 | < 0.001 | 0.441 | < 0.001 | < 0.001 |
Lesion conspicuity | 3.67 ± 0.52 | 3.67 ± 0.54 | 3.42 ± 0.36 | < 0.001 | > 0.999 | < 0.001 | < 0.001 |
Overall image quality | |||||||
b1000 | 3.50 ± 0.44 | 3.65 ± 0.36 | 3.06 ± 0.46 | < 0.001 | 0.054 | < 0.001 | < 0.001 |
ADC | 3.49 ± 0.45 | 3.50 ± 0.39 | 3.13 ± 0.40 | < 0.001 | > 0.999 | < 0.001 | < 0.001 |
Means and standard deviations of image quality obtained by two radiologists are summarized. Sharpness, artifact, and lesion conspicuity were evaluated in DWI (b1000). *p values were obtained from overall comparison among C-DWI, SMS2-DWI, and SMS3-DWI, †Dunn-Bonferroni post-hoc test was performed in case of significant differences in overall Friedman test. b1000, b = 1000 s/mm2. ADC = apparent diffusion coefficient