Table 1. Imaging parameters for reference and 3D multi-contrast sequences.

Parameters Reference 3D
T1-TSE T2-TSE 3D-MPRAGE 3D-MERGE T2-VISTA SNAP
FOV (mm3) 140×140×32 160×160×48 250×160×40
Resolution (mm3) 0.6×0.6×2 0.8×0.8×0.8
TE (ms) 10 50 5.3 4.2 255/86 4.5
TR (ms) 800 4,800 8.8/609 9 2,500 10/1,987§
TI (ms) N/A N/A 304 N/A N/A N/A
Flip angle (deg) 90 90 15 6 Variable 11/5
Turbo factor 10 12 32 90 130 + 4# 98
Signal acquisitions 1 2 1 2
Scan time 6:11 3:50 2:28 3:10 4:32 6:47

, Effective echo time/Equivalent echo time. , TR for one inversion cycle. §, SPGR repetition time/inversion recovery repetition time (IRTR). , flip angle used in IR/reference acquisition. #, Number of echoes in each shot for acquisition start-up. T1-TSE, T1-weighted turbo spin echo; T2-TSE, T2-weighted turbo spin echo; 3D-MPRAGE, three-dimensional magnetization-prepared rapid acquisition gradient-echo; 3D-MERGE, 3D motion-sensitized driven equilibrium prepared rapid gradient echo; T2-VISTA, T2-weighted volumetric isotropic turbo spin echo acquisition; SNAP, simultaneous non-contrast angiography and intraplaque hemorrhage; FOV, field of view; TE, echo time; TR, repetition time; TI, inversion time.