Skip to main content
. 2015 Dec 24;43(1):262–267. doi: 10.1118/1.4938266

Table II.

Automatching results between UTE‐MRI DRR and orthogonal kV imaging pair (LLAT and AP) for all seven patients.

Calculated shifts (cm) Known shifts (cm) Error (cm)
Patient no. Vertical Longi‐tudinal Lateral Vertical Longi‐tudinal Lateral Vertical Longi‐tudinal Lateral
Pt. 1 1.0 −1.5 1.0 1.0 −1.5 1.0 0.0 0.0 0.0
Pt. 2 0.8 −0.9 0.5 0.7 −1.0 0.5 0.1 0.1 0.0
Pt. 3 5.0 0.5 3.0 5.0 0.5 3.0 0.0 0.0 0.0
Pt. 4 0.5 0.5 1.0 0.5 0.5 1.0 0.0 0.0 0.0
Pt. 5 −0.1 0.3 0.1 −0.1 0.3 0.1 0.0 0.0 0.0
Pt. 6 −0.2 0.4 0.4 −0.2 0.4 0.4 0.0 0.0 0.0
Pt. 7 −2.0 −0.9 2.0 −2.0 −0.8 2.0 0.0 −0.1 0.0