. 2020 Jan 9;6:309. doi: 10.3389/fmed.2019.00309

Table 2.

Proportion of fractured tubes upon freezing in Biostore and Sample Store I at different fill volumes of distilled water.

Fill volume (μl) Brand 1 (HD) Brand 2 (HD) Brand 3 (HD) Brand 1 (SD) Brand 1 (HD—external)
Biostore 900 0.0% 66.7% 83.3% 8.3% 0.0%
970 70.8% 33.3% 50.0% 45.8% 0.0%
1,000 100.0% 66.7% 83.3% 95.8% 0.0%
Sample Store 900 37.5% 0.0% 0.0% 0.0% 0.0%
970 16.7% 0.0% 0.0% 0.0% 0.0%
1,000 41.7% 0.0% 0.0% 100.0% 0.0%

HD, samples frozen seated in high density tray; SD, samples frozen seated in standard density SBS racks.