Table 4.

Temperature homogeneity for sample store and biostore.

Sample Store Biostore
Average temperature (24 h) −22.19 ± 0.77 −79.84 ± 1.99
Max temperature (24 h) −17.19 −74.90
Min temperature (24 h) −25.03 −84.10
Peak variance (24 h) 7.84 9.20
Stability (24 h) 4.67 1.13
Uniformity (24 h) 2.32 8.07
Stability (DOC1) 1.69 0.87
Uniformity (DOC1) 2.29 8.07
Stability (DOC2) 1.78 0.28
Uniformity (DOC2) 2.02 8.06
Warm spot Row 3, column 1, front Row 5, column 1, front
Cold spot Row 38, column 30, mid Row 5, column 13, mid

DOC, door opening an closing, calculated for a 3 h period starting at the first door opening exercise (DOC1) and calculated for a 3 h period starting 3 h after the last door closing event (DOC2).