TABLE 2. Final Device Bench Testing Using Entire System of Devices and Software.
| Events | Air | Calibrat | Valsalva | Rest 1 | Kege Hold | Rest 2 | Kege Rep | Rest 3 | Air 2 | ||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Mass (g) | 0 | Spring Clamp | 200 | 0 | 500 | 0 | 1000 | 0 | 0 | ||
| Force (N) | 0 | 10.40 | 1.96 | 0 | 4.90 | 0 | 9.81 | 0 | 0 | ||
| Speculum Reponse (N) | IVT Number | 715 | −0.02 | 10.62 | 2.00 | 0.37 | 5.13 | 0.37 | 9.93 | 0.18 | 0.02 |
| 705 | −0.01 | 10.32 | 2.05 | 0.36 | 5.14 | 0.05 | 9.81 | 0.15 | 0.12 | ||
| 714 | −0.01 | 10.36 | 2.17 | −0.04 | 5.24 | 0.02 | 9.94 | 0.03 | 0.00 | ||
| 711 | −0.03 | 10.31 | 2.09 | 0.03 | 5.63 | 0.04 | 10.16 | 0.07 | 0.01 | ||
| 707 | −0.02 | 10.36 | 1.75 | 0.03 | 4.83 | 0.08 | 9.93 | 0.05 | 0.03 | ||
| 708 | 0 | 10.30 | 2.15 | 0.03 | 5.07 | −0.01 | 9.65 | 0.20 | 0.11 | ||
| Average | −0.02 | 10.38 | 2.04 | 0.13 | 5.17 | 0.09 | 9.90 | 0.11 | 0.05 | ||
| Std. Dev. | 0.01 | 0.12 | 0.15 | 0.18 | 0.26 | 0.14 | 0.17 | 0.07 | 0.05 | ||