Skip to main content
. 2020 Jan 2;15(1):1–8. doi: 10.4103/atm.ATM_24_19

Table 2.

Proposed solutions for some of the causes of patient-ventilator asynchrony

Cause Description Solution
Auto-triggering It is defined as the occurrence of at least three consecutive pressurizations at a ventilator frequency[8] of >40/min not synchronized with patient respiration
The trigger is too sensitive or nonrespiratory factors trigger the ventilator such as
 Cardiac contractions: This may cause a small amount of air movement, and if the flow trigger is sensitive enough, this air movement can trigger the ventilator breaths. The respiratory rate may match the heart rate
 Leak from the circuit or from the chest drain (e.g., a bronchopleural fistula)
 Inappropriate sensitivity settings
 Excessive water condensation in the ventilator circuit
 Large volume of respiratory secretions
Swallowing or vomiting
 Peristalsis in a massive hiatus hernia or intrathoracic bowel loops
 Muscle contractions due to external pacing
Remove the cause such as excessive water condensation or respiratory secretions
If no treatable cause is detected, adjust the trigger to a higher setting
Double-triggering When an insufficient level of pressure support is applied or the patient’s demand is high, the inspiratory effort may continue throughout the preset ventilator inspiratory time and result in retriggering of the ventilator after it has discontinued pressurization, which may lead to the delivery of two cycles for only one patient’s effort (double-triggering) Adjust the expiratory flow trigger until the desired tidal volume is achieved
Large leak around the NIV mask In order to generate the specified pressure, the ventilator continues to deliver high flow. With a large leak, inspiration can be very uncomfortable (as the ventilator delivers 70-80 L/min of gas into the patients face)[5] Adjust the mask to minimize the leak
Decreasing the level of pressure support will decrease the total inspiratory time, as the machine will cycle to expiration sooner
In some ventilators, one can actually adjust the inspiratory time directly
Wasted effort Wasted efforts can occur during inspiration when the patient tries to initiate a breath (straining to inhale against a closed inspiratory valve)
Possible causes
 Respiratory muscle weakness
 Reduced respiratory drive
 Inadequate trigger threshold setting
 iPEEP: Inspiratory muscles have to overcome iPEEP and trigger sensitivity to trigger the ventilator
 Inadequate level of support: The flow rate is too low and it does not meet patient demand
The use of ePEEP to approximately 80%-90% of the iPEEP can counterbalance iPEEP and as such facilitate triggering
Reduce the trigger sensitivity

PEEP=Positive end-expiratory pressure, ePEEP=Extrinsic PEEP, iPEEP=Intrinsic-PEEP, NIV=Noninvasive ventilation